Siemens Medical Solutions E.CAM Patient Handling System (PHS) Field of View (FOV) indicator bars; a component of the E.C
FDA device recall Z-1505-2009 · posted 2009-06-24 · Terminated
Recall details
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Reason for recall
- Patient fingers can get pinched if they attempt to pull themselves up from the patient handling system without any assistance and grabbing onto the flip-up FOV indicator bars.
- Action taken
- Siemens sent Urgent Field Correction Recall letters dated 12/18/08 to all users of the E.cam Systems. The accounts were informed of the potential risk to patients when they attempt to pull themselves up from the patient handling system (PHS) without any assistance and grabbing the field of view (FOV) indicator bar (aka flip-up armrests). The customers were advised to always assist the patient when getting on/off of the PHS. They also were advised that a Siemens Customer Service Representative would contact them within 90 days to schedule a corrective upgrade to the PHS with a modified FOV guide to mitigate the risk for patient injury free of charge. Any questions were directed to contact Siemens Medical Solutions USA, Inc. at 1-800-767-2313 (USA) for assistance.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- KPS
- Quantity affected
- 1,107 units
- Distribution
- Worldwide distribution: USA, Guam, Argentina, Australia, Austria, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Czech Republic, Ecuador, Egypt, Finland, France, Germany, G
- Date initiated
- 2008-12-18
- Date posted
- 2009-06-24
- Date terminated
- 2011-05-03
- Location
- Hoffman Estates, IL