Siemens Healthcare Diagnostics Vista Air Compressor. Vista AirCompressor replacement part: SMN 10706557 SKIT Air Compres

FDA device recall Z-1733-2014 · posted 2014-06-09 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Reason for recall
A small number of Vista Air Compressors were miswired, which if installed and powered up, could cause a fire, or become an electrical shock hazard.
Action taken
Siemens sent an Urgent Medical Device Correction letter dated April 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Actions to be Taken by the Customer 1. If an Air Compressor has already been delivered to your laboratory, or is delivered in the future for installation by your Siemens Healthcare Customer Service Engineer, please do not make any attempt to install the Air Compressor by yourself. 2. Please notify your local Siemens Healthcare Diagnostics Customer Care Center when an Air Compressor is at your site so arrangements may be made for installation by a Siemens Healthcare Diagnostics Customer Service Engineer. Please discuss the content of this letter with your Medical Director. Please complete and return the Field Correction Effectiveness Check Form attached to this letter. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. We apologize for the inconvenience this situation has caused. If you have any questions, please contact Siemens Customer Care Center - Technical Solutions at (800-441-9250) or your local Siemens technical support representative.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JJE
Quantity affected
4
Distribution
Worldwide Distribution: US Distribution to states of: IL, OH and TN; and country of: Italy.
Date initiated
2014-04-14
Date posted
2014-06-09
Date terminated
2014-12-15
Location
Newark, DE