Siemens Healthcare Diagnostics, Inc., Dimension Vista Intelligent Laboratory System with Alpha Pinnacle 10kVA UPS Extern
FDA device recall Z-2423-2010 · posted 2010-09-15 · Terminated
Recall details
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Reason for recall
- Defective batteries used in the Alpha Pinnacle P10 UPS External Battery Pack may overheat and cause a strong sulfuric odor to be released and a corrosive sulfuric acid solution leak to occur.
- Action taken
- Siemens issued an Urgent Field Safety Notice in May 2010 to customers with information on what to do if the battery leaks, overheats, or vents an odor. Firm also issued a Support Bulletin to field service personnel to inspect and replace affected units.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- JJE
- Quantity affected
- 228 units
- Distribution
- Worldwide distribution: USA, Australia, Austria, Belgium, Canada, France, Germany, Italy, Japan, Malaysia, Netherlands, New Zealand, Portugal, Slovenia, Spain, and Switzerland.
- Date initiated
- 2010-05-05
- Date posted
- 2010-09-15
- Date terminated
- 2010-10-25
- Location
- Newark, DE