Siemens Healthcare Caps for capillary 140/175 uL, Package count 100 For use with blood collection Multicap glass and Mul

FDA device recall Z-1257-2015 · posted 2015-03-06 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics Inc
Reason for recall
Capillary Caps for RAPIDLyte Multicap and Multicap-S Blood Collection Capillaries may not attach and result in breakage of the glass capillary
Action taken
Siemens Healthcare issued an Urgent Field Safety Notice via e-mail to affected Siemens Healthcare Diagnostics Regional offices on January 22, 2015 both in the United States and elsewhere for communication with affected customers. Siemens advises customers to use caution when applying the Capillary Caps with Siemens Multicap glass and Multicap-S plastic capillaries to minimize the potential for breakage or bending.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JKA
Quantity affected
32384
Distribution
Worldwide Distribution. US Nationwide, Australia, Austria, Belgium, Bosnia, Herzegovina, Canada, Colombia, Croatia, Curacao, Czech Republic, Estonia, Finland, France, Georgia, Germany, Great Britain,
Date initiated
2015-01-22
Date posted
2015-03-06
Date terminated
2016-04-13
Location
Norwood, MA