siemens Dimension Phenytoin Flex Reagent Cartridge The PTN method used on the Dimension clinical chemistry system is an
FDA device recall Z-0148-2015 · posted 2014-10-29 · Terminated
Recall details
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Reason for recall
- Internal testing has confirmed a low absorbance range between the levels 1 and 5 calibrators with this lot resulting in the potential for a positive or negative bias and imprecision across the PTN assay range.
- Action taken
- Siemens sent an Urgent Medical Device Recall Notification dated September 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed as follows: " Discontinue use and discard any remaining inventory of PTN lot FA5154, and complete the attached form indicating your replacement product needs in the attached Field Correction Effectiveness Form. " Siemens will replace any unused inventory of the affected lot at no charge. We apologize for the inconvenience this situation may have cause. If you have any questions, please contact Siemens Customer Care Center - Technical Solutions at (800) 441-9250, or your local Siemens technical support representative.
- Root cause
- Under Investigation by firm
- Status
- Terminated
- Product code
- DIP
- Quantity affected
- 1733
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico., and the countries of Australia, Brazil, Bahamas, Canada, Chile, India, Japan and Uruguay
- Date initiated
- 2014-09-24
- Date posted
- 2014-10-29
- Date terminated
- 2016-08-17
- Location
- Newark, DE
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| DU PONT DIMENSION(R) PHENYTOIN (PTN) METHOD (K911056) | E.I. Dupont DE Nemours & Co., Inc. | 1991-04-03 |