Atlas FDA

Siemens Dimension LOCI TNI Troponin I Calibrator RC 621. The LOCI TNI Troponin I Calibrator is an in vitro diagnostic pr

FDA device recall Z-1926-2014 · posted 2014-06-26 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Reason for recall
Siemens Dimension LOCI TNI Troponin I calibrator lot 3DD007 may demonstrate an average upward shift of 24% in QC and patient results following calibration.
Action taken
Siemens sent an Urgent Medical Device Recall letter dated September 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were advised immediate discontinuation of use and discard of the affected device. Please complete the attached Effectiveness Check form and fax it to the Siemens Customer Care Center at (302) 631-8467 to indicate that you have received this information. Also, please indicate any replacement product required. Please contact Siemens Customer Care Center- Technical Solutions at (800) 441 9250 if you are aware of any adverse event that occurred as a consequence of this issue, or for immediate product needs. Please retain this letter with your laboratory records, and forward this Jetter to those who may have received this product. We apologize for the inconvenience this situation has caused. If you have any questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative.
Root cause
Process control
Status
Terminated
Product code
JIT
Quantity affected
1625
Distribution
Distribution: US Nationwide including Puerto Rico
Date initiated
2013-09-17
Date posted
2014-06-26
Date terminated
2014-08-27
Location
Newark, DE