Siemens Cios Alpha VA30-mobile X-Ray system, Material # 11105200 The Cios Alpha and Cios Spin are a mobile X-Ray system

FDA device recall Z-1958-2020 · posted 2020-05-04 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector
Action taken
Siemens Healthcare issued A Customer Safety Advisory Notice dated 4/20/20 via AX030/20/S. Letter states reason for recall, health risk and action to take: We strongly recommend that you do not disconnect the main unit from the monitor trolley when the system displays the system error message Err 16305 / 80: Confirm this error and repeat your last action until the problem has been corrected by a service engineer. Contact our service organization at 1-800-888-7436. Siemens is currently developing a solution to eliminate the root cause of this problem.
Root cause
Process design
Status
Terminated
Product code
OWB
Quantity affected
16 units
Distribution
US Nationwide distributions.
Date initiated
2020-04-20
Date posted
2020-05-04
Date terminated
2021-05-10
Location
Malvern, PA