Siemens brand 550 TxT Patient Table, Part No. 7346534, Distributed by and/or Manufactured by Siemens AG, Medical Solutio

FDA device recall Z-0166-2011 · posted 2010-10-28 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
An increase in voltage may cause a short circuit of a power transistor that may cause a smoldering and smoking effect of the electrical wiring and also an unexpected lowering of the Table. Both issues present the potential for injury to a patient or the operator.
Action taken
Siemens sent a Customer Information Letter on March 31, 2010, to all consignees. The letter identified the product, the problem, and the action that should be taken. Siemens implemented minor modifications to address the issues. No further action was required for the continual operation of the 550TxT Table. For questions regarding this recall call 925-602-8083.
Root cause
Other
Status
Terminated
Product code
JAI
Quantity affected
326 units
Distribution
Worldwide Distribution including USA
Date initiated
2010-03-31
Date posted
2010-10-28
Date terminated
2012-03-01
Location
Concord, CA