Si emans Simview NT Cassette Holder Electronic Imaging Device (Image Identifier) with CT, Part No: 54 96 984; Siemens Me

FDA device recall Z-0283-2008 · posted 2008-01-31 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
Loose Mounting Screws -- Cassette holder may become loose and result in the device dislodging and colliding with the patient.
Action taken
Customers were sent a Customer Safety Advisory Notice on February 27, 2007, notifying them that the mounting screws on the cassette holders on the SIMVIEW 3000 and NT Simulators, as well as some MEVASIM Simulators, could loosen and can cause damage to the system, or personal injury. The firm requested that they check of tightness on the mounting hardware. A STOP Modification notice was issued on 06/01/2007, to all affected regional service organizations that are responsible for implementing updates due to new information. Updated letters and User instructions will be released in August 2007.
Root cause
Labeling design
Status
Terminated
Product code
KPQ
Quantity affected
287 units total, all recalls Z-0281-285-2008
Distribution
Worldwide Distribution - USA including states of IN, VA, MA, VT, NJ, NY, PA, DC, NC, SC, GA, AL, TN, MI, KY, OH, IN MI, MN, ND, MT, IL, MO, KS, NE, AR, OK, TX, AZ, CA, SD, LA, Puerto Rico, and countri
Date initiated
2007-02-27
Date posted
2008-01-31
Date terminated
2011-02-03
Location
Concord, CA