Short Percutaneous Antenna with Thermosphere" Technology Standard Percutaneous Antenna with Thermosphere" Technology; Th

FDA device recall Z-1187-2016 · posted 2016-03-17 · Terminated

Recall details

Recalling firm
Covidien Medical Products (Shanghai) Manufacturing L.L.C.
Reason for recall
Medtronic is voluntarily recalling specific item codes and production lots of Covidien Emprint" Percutaneous Antenna with Thermosphere" Technology due to the potential for the generator to shut down in the precence of high temperature measurements in the Emprint" Percutaneous Antenna.
Action taken
Medtronic sent an Urgent Medical Device Recall letter dated February 23, 2016 to all affected customers via Federal Express or Certified Mail. The letter identified the affected product, problem and actions. Customers were advised to quarantine and discontinue use of the affected product and to Return product. For questions or concerns, contact your Medtronic representative or Customer Service at (800) 882-5878.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
NEY
Quantity affected
4,920 units
Distribution
Worldwide Distribution - US Nationwide, VA/govt/military consignees.
Date initiated
2016-02-23
Date posted
2016-03-17
Date terminated
2017-06-21