Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shave

FDA device recall Z-1872-2014 · posted 2014-06-25 · Terminated

Recall details

Recalling firm
Stryker Endoscopy
Reason for recall
STERRAD 100S parameters provided in the Shaver Handpiece reprocessing guide (P10299 Revision A) did not fully consider all worst-case scenarios during validation testing.
Action taken
Device Correction letters were sent on May 23, 2014 to all users by certified mail. Internation sites will receive notification and be responsible for contacting customers.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
HRX
Quantity affected
1992 trays
Distribution
Worldwide Distribution.
Date initiated
2014-05-19
Date posted
2014-06-25
Date terminated
2015-05-07
Location
San Jose, CA