SHAPE ARM, SINGLE, an articulating, arm bracket that is attached to the service head of boom systems which are used for

FDA device recall Z-2052-2014 · posted 2014-07-15 · Terminated

Recall details

Recalling firm
Stryker Communications
Reason for recall
During the use of the SHAPE Arm there is the potential for the monitor mount to come off of the SHAPE Arm if the lock washer and screw are not present.
Action taken
Direct accounts in the US were notified by letter dated March 21, 2014, and asked to report any malfunctions to Stryker Communications immediately. In addition, Stryker Representatives will visit all US accounts to inspect all the product units on hand. If needed, Representatives will check that black, screw/lock washers supplied by Stryker are secured in place and will replace any that are not the typed furnished by Stryker. All units will be updated with a yellow, warning sticker on the monitor bracket which reads: "Failure to properly install the screw and lock washer below can result in the attached component/equipment falling." Stryker Communications' foreign distributors will oversee notification, inspection, and corrections of their accounts and report corrections and other findings to Stryker Communications.
Root cause
Process control
Status
Terminated
Product code
BRY
Quantity affected
953 STRYKER SHAPE ARMS of all types
Distribution
Worldwide Distribution. US nationwide including Puerto Rico, China, Philippines, Mexico, Korea, Japan, Colombia, Chile, Canada, Brazil, and Australia
Date initiated
2014-03-17
Date posted
2014-07-15
Date terminated
2015-05-15
Location
Flower Mound, TX