Shandon Rapid Chrome Kwik-Diff Kit Part Number: 9990700 Shandon Rapid Chrome Kwik-Diff Reagent #3 Part Number 9990707 Sh
FDA device recall Z-1895-2017 · posted 2017-04-17 · Terminated
Recall details
- Recalling firm
- Richard-Allan Scientific Company
- Reason for recall
- Methylene Blue, Solution #3 in the Kwik Diff Kit was not Methylene Blue but Crystal Violet instead.
- Action taken
- On 3/15/2017, URGENT MEDICAL DEVICE RECALL letters were sent to the affected consignees. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. ACTIONS TO BE TAKEN BY THE CUSTOMER/USER Please contact Richard Allan Scientific for directions on safe disposal or return if customers have affected lots. All other lots of the products, or similar products manufactured by Richard Allan Scientific, are unaffected. Upon identification of the affected lot, please notify Richard Allan Scientific of the quantity and the preferred disposal/return method. If product is being disposed of, evidence of disposal is required to be sent to Richard Allan Scientific. If affected product has been used with no issues, please notify Richard Allan Scientific of the quantity that is unavailable for return or disposal. Please use the attached Recall Return Response Acknowledgement & Receipt Form.
- Root cause
- Mixed-up of materials/components
- Status
- Terminated
- Product code
- HYF
- Quantity affected
- 390
- Distribution
- Worldwide Distribution -- US: CA, CO, CT, DE, GA, HI, IA, IL, KS, MA, MS, NC, NH, NM, NY, OH, PA, PR, RI, UT, VA, VT, WA, and WV; and, the countries of AE, BH, CA, CZ, FR, GB, HK, SA, and TR.
- Date initiated
- 2017-03-15
- Date posted
- 2017-04-17
- Date terminated
- 2017-07-27
- Location
- Kalamazoo, MI