seprafilm Single Site ADHESION BARRIER Re-order Number: 6641-01 Product Usage: Seprafilm Adhesion Barrier is indicated f
FDA device recall Z-1856-2012 · posted 2012-06-21 · Terminated
Recall details
- Recalling firm
- Genzyme Corporation
- Reason for recall
- Sterility of product may be compromised due to packaging defect
- Action taken
- Genzyme issued an Urgent: Medicatical Device Product Recall Notification Letter dated June 13, 2012 to all affected customers. Stericycle Expert Solutions, will be acting on behalf of Genzyme to coordinate notifications, communications, and logistics for returns of all product subject to this recall. The letter identified the affected products, problen and actions to be taken. Customers are requested to check their inventory and return all affected products per the instructions provided. Customers are instructed to complete the enclosed Business Reply Form and return via fax transmission to 866-912-2519. For questions call Stericycle at: 877- 473-7069.
- Root cause
- Packaging process control
- Status
- Terminated
- Product code
- MCN
- Quantity affected
- 1285 units
- Distribution
- Worldwide Distribution - US (nationwide) and the country of S. Korea
- Date initiated
- 2012-06-13
- Date posted
- 2012-06-21
- Date terminated
- 2013-07-24
- Location
- Framingham, MA
Monitor recalls programmatically