Selenia Dimensions Mammography System, Model number ASY-04160, Selenia Dimensions Gantry, UDI 15420045505636,

FDA device recall Z-1343-2020 · posted 2020-02-18 · Terminated

Recall details

Recalling firm
Hologic, Inc.
Reason for recall
The recalling firm has received reports from the field of the C-arm on the Selenia Dimensions system seizing or dropping during field repairs and cleaning of the systems.
Action taken
On January 21, 2020, the firm notified affected customers of the recall via hand-delivered Urgent Medical Device Correction letter. Customers were informed that the firm had received reports of the C-arm seizing or dropping during field repairs and cleaning of the systems. In order to ensure continued safe use of the system, Hologic sent field service engineers to customer sites on January 21, 2020, to inspect affected systems. Customers were asked to not use the system until the inspection was completed. If you have any questions, please contact Technical Support at breasthealth.support@hologic.com
Root cause
Process control
Status
Terminated
Product code
MUE
Quantity affected
72
Distribution
Domestic distribution to 14 states: AK AZ CA CO CT DC FL GA IA ID IL MA MO NC NH NJ NY OH SD TX VA. Foreign distribution to Italy, United Kingdom, Saudi Arabia, Brazil, Costa Rica, Spain, Canada, Cypr
Date initiated
2020-01-21
Date posted
2020-02-18
Date terminated
2021-08-11
Location
Danbury, CT