Atlas FDA

Segmental Stem, Cemented, Bowed, 152MM, 1 EACH. Model Numbers: CB-11152-03M, CB-13152-03M, CB-15152-03M, CB-17152-03M. C

FDA device recall Z-1880-2025 · posted 2025-05-29 · Open, Classified

Recall details

Recalling firm
Onkos Surgical, Inc.
Reason for recall
Potential for breach of Tyvek seals.
Action taken
Onkos Surgical began notifying consignees on about 09/01/2023 via email. Consignees were instructed to complete and return the response form providing the affected lot numbers being returned, coordinate the return of affected units, and to note that when placing these parts into totes, always place them flat.
Root cause
Packaging change control
Status
Open, Classified
Product code
KRO
Quantity affected
49 units
Distribution
Product was shipped from Onkos' distribution in Olive Branch, MS to consignees in the following states: GA, TX, WA, LA, NJ, FL, AZ, MD, MS, KS, IL, WI, KY, PA, NC, NE, OH, HI.
Date initiated
2023-09-01
Date posted
2025-05-29
Location
Parsippany, NJ

Monitor recalls programmatically