Atlas FDA

Sechrist Stryker Medical Stretchers Model 742 Intended to transport patients in a horizontal position.

FDA device recall Z-2861-2011 · posted 2011-07-22 · Terminated

Recall details

Recalling firm
Stryker Medical Division of Stryker Corporation
Reason for recall
The firm has determined that the brake rod assemblies of various stretchers that were built with a 5th wheel were manufactured with roll pins that did not meet the materials specifications. Some roll pins may have fractured during the assembly process and could fall out of the brake rod assemblies, rendering the brake or steer systems inoperable from certain pedals or causing a reduction in overa
Action taken
Stryker Medical sent an "Urgent Medical Device Correction" dated July 19 ,2011 to all affected customers. The letter included; the name and model numbers of the stretchers, the reason for the correction and dates of manufacture. It contained the instructions to check the braking system and if found to be inoperable, to remove it from service, and if operable to check prior to each use. Also, information regarding contacting the recipients of any loaned or sold stretchers to relay the information in the letter and Medwatch reporting information. A self addressed postcard to confirm receipt of the letter was also included. For additional information call (269) 389-6604.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
FPO
Quantity affected
10
Distribution
Worldwide Distribution: Nationwide (USA) including the states of NY, MI, LA and CA ,and the countries of Canada, China, Spain, Japan, Korea, United Kingdom, South Africa.
Date initiated
2011-07-19
Date posted
2011-07-22
Date terminated
2013-08-29
Location
Portage, MI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.