SEAC Advanced Dental System, Model A591561 The SEAC Advanced Dental System is a self-contained dental delivery system wi

FDA device recall Z-0545-2014 · posted 2013-12-19 · Terminated

Recall details

Recalling firm
ASI Medical, Inc.
Reason for recall
ASI Medical is initiating a field correction for the SEAC Advanced Dental System because the miniature plastic air tank can potentially rupture.
Action taken
ASI Medical sent an Field Correction Action and Safety Notice letter dated November 5, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. REQUIRED FIELD CORRECTION Our field correction will consist of sending you a warning label to be applied to the back service access panel of the cart and an ASME safety pop off valve that will open and exhaust if pressure should reach over 100 psi. This safety pop off valve is designed be easily installed in line between the air compressor and air tank in your system. We are requesting your cooperation in making this field correction.
Root cause
Device Design
Status
Terminated
Product code
EIA
Quantity affected
337
Distribution
Worldwide Distribution - USA (nationwide) including the countries of Austria, Aruba, Saudi Arabia, Canada, Germany, Spain and West Indies.
Date initiated
2013-11-05
Date posted
2013-12-19
Date terminated
2014-08-11
Location
Centennial, CO