Screwdriver Blade, Handle, Drill Bit, Part #313.945 Low Profile Neuro Screwdriver Blade/MR Safe/MQC/32mm Short Lot # 505
FDA device recall Z-2189-2016 · posted 2016-07-12 · Terminated
Recall details
- Recalling firm
- Synthes (USA) Products LLC
- Reason for recall
- DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
- Action taken
- DePuy Synthes mailed an URGENT MEDICAL DEVICE FIELD SAFETY NOTIFICATION letter to their customers (Hospital Personnel, including Imaging Department Personnel and DePuy Synthes Colleague) to inform them of the recall and actions to take. The letter identified the affected, problem and actions to be taken. The letter instructed customers not to use the affected Instruments which are labeled or etched MR Safe and MR Conditional during MRI scans. Customers were asked to review the information contained in the Field Safety Notification letter and complete the Verification Section located on page 3 of this notification. For questions call 610-719-5450 or contact your DePuy Synthes Sales Consultant.
- Root cause
- Labeling False and Misleading
- Status
- Terminated
- Product code
- HWE
- Quantity affected
- 50 units
- Distribution
- US Nationwide Distribution in the states of: AR,CA,CO,CT,DC,ID,IL,KY,LA,MA,MD,MN,MO,MS,MT,NY,PA and TX.
- Date initiated
- 2016-05-31
- Date posted
- 2016-07-12
- Date terminated
- 2017-02-02
- Location
- West Chester, PA