Sarns " TCM II Temperature Control Module The Sarns" TCM II Temperature Control Module is a source of temperature-contro

FDA device recall Z-1569-2013 · posted 2013-06-21 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
Internal testing found certain Sarns" TCM II systems were released with incorrectly assembled heatsink assemblies which could cause the units to fail.
Action taken
Terumo sent beginning May 30, 2013, an URGENT MEDICAL DEVICE RECALL NOTIFICATION were sent to all customers with instructions to notify all customers of the recall requesting all affected units to be quarantined until the heatsink units are replaced Confirm receipt by filling out the response form containing communication by faxing or emailing the attached Affiliate Response Form to the fax number or email address indicated on the form. Questions should be directed to Terumo CVS customer service in the U.S.: Telephone: +1-734-663-4145; Fax: +1-734-741-6149
Root cause
Employee error
Status
Terminated
Product code
DWC
Quantity affected
7 units
Distribution
Worldwide Distribution including the countries of UAE (United Arab Emirates) HONG KONG and THAILAND. Not distributed in USA
Date initiated
2013-05-30
Date posted
2013-06-21
Date terminated
2013-09-17
Location
Ann Arbor, MI