Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Reason for recall
- During production of the Sarns Antegrade Cardioplegia Cannula, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of a foreign substance on the inner surface of some cannulae tips. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface. The substance was likely deposited during the molding process, but the exact composition of the
- Action taken
- Terumo sent an Urgent Medical Device Recall letter dated May 9, 2012 to All US first consignees will receive a removal notification letter via Federal Express. The letter identified the affected products, reason for correction and actions to be taken. Customers were instructed to discontinue use of affected products and return for replacement. The letter advised customers to assure that all users are aware of this notice and to fill out and return the attached Response Form. For questions call Terumo CVS Customer Service at 1-800-521-2818 or by fax at 1-800-292-6551.
- Root cause
- Material/Component Contamination
- Status
- Terminated
- Product code
- DWF
- Quantity affected
- 53,139 (each) in total
- Distribution
- Worldwide Distribution - US (nationwide) including the states of: AL, AR , AZ , CA, CT , FL, IA , IL, IN , KS , LA , MI , NM, MO , NC, NE , NJ , NV, NY , OH, OK , PA, RI , SC , TN, TX, WA , WI, WV and
- Date initiated
- 2012-05-09
- Date posted
- 2012-06-27
- Date terminated
- 2013-04-24
- Location
- Ann Arbor, MI
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| SARNS CARDIOPLEGIA/VENT CATHETERS (K873212) | 3M Health Care, Sarns | 1987-11-06 |