Atlas FDA

Sandel Healthcare Safety Devices Weighted Safety Scalpel #3 Disposable Safety Handle with #15 Blade Manufactured for: An

FDA device recall Z-0639-2014 · posted 2014-01-08 · Terminated

Recall details

Recalling firm
Ansell Sandel Medical Industries LLC
Reason for recall
Select lots of SANDEL Weighted Safety Scalpels are packaged with the scalpel blade exposed beyond the safety shield which could compromise the integrity of the package and the sterility of the scalpel leading to injury.
Action taken
The firm, Ansell, sent an "URGENT: VOLUNTARY RECALL OF ANSELL SANDEL WEIGHTED SAFET SCALPEL" letter dated November 11, 2013 to its customers and posted a website notification on November 11, 2013. The letter described the product, problem and actions to be taken. The customers were instructed: Do not ship SANDEL Weighted Safety Scalpel products 2210-L, 2210-N, 2211-L, 2211-N, 2215-L, 2215-N, or 2220-N to a hospital or end user with the listed lot numbers; quarantine SANDEL Weighted Safety Scalpel products in your inventory; and complete the attached response form and return it to attention: Ansell; via fax at 818-534-2510 or email sandel-cs@ansell.com. Additionally, if you have distributed this product to other accounts, please notify them of the voluntary recall and instruct them to return any outstanding stock to you as indicated in the attachment. If you have any questions or need additional information, please contact Customer Service by calling 866-764-3327 (6:00am-5:00pm PST/9:00am-8pm EST) or email sandel-cs@ansell.com
Root cause
Under Investigation by firm
Status
Terminated
Product code
GDZ
Quantity affected
1203 - 96 cases (4128 units) ; 1553 -96 cases (3548 units)
Distribution
Worldwide Distribution: US (nationwide) including states of: AL, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MN, MS, NC, NE, NY, OK, PA, PR, RI, TN, TX and WI; and country of: Canada.
Date initiated
2013-11-12
Date posted
2014-01-08
Date terminated
2016-04-06
Location
Chatsworth, CA

Monitor recalls programmatically