Safety PICC Insertion Tray with Chloraprep OneStep - Model #5026CL-9 - 12 pr case. Packaged for Physician use only.

FDA device recall Z-2096-2015 · posted 2015-07-17 · Terminated

Recall details

Recalling firm
Custom Healthcare Systems, Inc.
Reason for recall
An INCORRECT item number is packed within the OUTSIDE "PIGGYBACKED" PACKAGE. The interior sealed STERILE KIT is correct and NO adverse issues affect that part of the item. The NON-STERILE OUTSIDE "PIGGYBACKED" PACKAGE mistakenly contains ITEM #8881579121 - Syringe prefilled, Saline Flush, 09% Sodium Chloride and is a NONSTERILE PRODUCT that CANNOT BE DROPPED INTO THE STERILE FIELD.
Action taken
Custom Healthcare System sent an Urgent Product Recall letter dated November 5, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. PLEASE CHECK YOUR INVENTORY AND QUARANTINE ANY AFFECTED PRODUCT MATCHING THE DESCRIPTIONS ABOVE. Call 804 (421-5959) with regards to returning the affected products. Return Authorizations will be issued and product pickup and return will be scheduled.
Root cause
Mixed-up of materials/components
Status
Terminated
Product code
GCB
Quantity affected
3 Cs-Lot#92414 , 25 Cs-Lot#100214, 24 Cs-Lot#100914
Distribution
US Distribution to CT.
Date initiated
2014-11-05
Date posted
2015-07-17
Date terminated
2016-01-12
Location
Richmond, VA