S5 Single Roller Pump 150, Item Number: 10-80-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopul
FDA device recall Z-0271-2013 · posted 2012-11-08 · Terminated
Recall details
- Recalling firm
- Sorin Group USA, Inc.
- Reason for recall
- Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determi
- Action taken
- The firm, Sorin Group, sent an "URGENT FIELD SAFETY NOTICE" dated September 19, 2012 to all affected customers by certified mail. The notice described the product, problem and actions to be taken. The customers were instructed to follow the provided instructions for the continued safe use of the Perfusion systems until they have been serviced or replaced by Sorin Group. The Sorin Group Service Team will contact the customer to schedule servicing and replacement of the affected product. The customers were also instructed to complete and return the Customer Response Form via fax to 303-467-6502 or by email to yvonne.feyerherm@sorin.com; assure this notice is distributed to all personnel who need to be aware of this notice; and if they have transferred the affected products to a third party, pass this information with them as well as with Sorin Group Customer Service at 1-800-650-2623. For questions regarding this notice, contact Sorin Group USA Customer Service at 1-800-650-2623.
- Root cause
- Mixed-up of materials/components
- Status
- Terminated
- Product code
- DTQ
- Quantity affected
- 35 units
- Distribution
- Worldwide distribution: USA (nationwide) including states of: AZ, GA, MI, MS, NC, NY, PA, SC, and TX; and countries of: Austria, Australia, Belgium, Canada, People's Republic of China, Columbia, Denma
- Date initiated
- 2012-09-19
- Date posted
- 2012-11-08
- Date terminated
- 2013-01-18
- Location
- Arvada, CO
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| SORIN C5 SYSTEM MODEL 58-00-00 (K093882) | Sorin Group Deutschland GmbH | 2010-01-15 |
| MODIFICATION TO: STOCKERT S5 SYSTEM (K071318) | Sorin Group Deutschland GmbH | 2007-07-06 |