RX ACCULINK Carotid Stent System 9 X 20mm, Part Number: 1011341-20, Lot Number: 0071361 The RX Acculink Carotid Stent Sy

FDA device recall Z-0337-2011 · posted 2010-11-10 · Terminated

Recall details

Recalling firm
Abbott Vascular
Reason for recall
The recall was initiated because Abbott vascular has discovered through internal testing that the affected lots may not meet the firm's quality specifications for catheter shaft tensile strength. If the affected device is used it may result in acute stent deployment difficulties and subsequent intervention.
Action taken
Abbott Vascular sent an URGENT DEVICE RECALL letter dated September 10, 2010, to all consignees. The letter identified the product, the problem, and the action to be taken. The Sales Representatives were to personally visit each account to deliver the recall letter and assist in identifying and returning any unused devices to Abbott Vascular. Also, Abbott Vascular Regulatory Compliance initiated contact with the Risk Management department at each account by phone to verify their mailing address and provide a courtesy copy of the recall letter. The consignees were instructed to work with their local account representative to review their inventory, complete the Recall Effectiveness Check and return identified products to Abbott Vascular. For questions regarding this recall call (951) 914-3324.
Root cause
Pending
Status
Terminated
Product code
NIM
Quantity affected
42 units total for all products
Distribution
Nationwide Distribution including AL, CA, CT, DC, FL, GA, IA, IL, IN, MA, ME, MI, MN, MO, OH, OK PA, TX, VA, and WI
Date initiated
2010-09-10
Date posted
2010-11-10
Date terminated
2010-11-16
Location
Temecula, CA