RSP Humeral Socket Reamer (Size 32 SML): DJO Surgical Instruments and Instrument Cases DJO surgical Encore Medical, L.P.

FDA device recall Z-0150-2012 · posted 2011-11-02 · Terminated

Recall details

Recalling firm
Encore Medical, Lp
Reason for recall
DJO Surgical has determined that the alignment pins cannot be consistently cleaned using the technique provided in the cleaning/sterilization section of the instrument IFU.
Action taken
DJO Surgical sent an URGENT FIELD SAFETY NOTICE letter dated September 28, 2011, to all affected customers. The letter identified the product. the problem, and the action to be taken by the customer. Customers were instructed to return the RSP Humeral Socket Reamers and RSP Baseplate Rim Planers. The RSP Humeral Socket Reamers will be reworked and returned to the customer. The Baseplast Rim Planer will be replaced when a suitable design is available. During the interim until instruments are returned, the use of FMP acetabular reamers size 42mm, 43mm, 45mm (DJO PIN 803-05-226, 227, 200) for the humeral preparation was validated to be effective to prepare the humerus for the implant. Customers were instructed to immediately notify the hospitals and surgeons using these devices and document it on the Field Safety Notice Response Form. For questions regarding this recall call 512-832-9500.
Root cause
Device Design
Status
Terminated
Product code
LXH
Quantity affected
399 Reamers
Distribution
Worldwide Distribution - USA including AR, AZ, CA, CO, FL,GA, ID, IN, IL, LA, MA, MD, ME, MI, MO, NJ, NM, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, and WI and Internationally toGermany, South Americ
Date initiated
2011-09-28
Date posted
2011-11-02
Date terminated
2013-03-29
Location
Austin, TX