Roller Pump, 6 Inch Diameter The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extrac
FDA device recall Z-0840-2012 · posted 2012-01-24 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Reason for recall
- Terumo Cardiovascular Systems (Terumo CVS) discovered and rejected a single digital encoder cable assembly during incoming inspection because it had a loose wire. The assembly is used in the 4" and the 6" roller pumps for the Terumo Advanced Perfusion System 1.
- Action taken
- TERUMO sent an Urgent Medical Device Correction letter dated January 13, 2012, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to review the Medical Device Correction notice. Assure that all users are aware of the notice. Confirm receipt of the communication by faxing the attached Customer Response Form to 1-900-292-6551. Roller pumps with affected cables will have the cables assembly replaced. For questions regarding this recall call 1-800-521-2818.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- DTQ
- Quantity affected
- 44
- Distribution
- Worldwide Distribution - USA including MD, WI, and IA and the countries of Canada, Mexico, and Vietnam
- Date initiated
- 2012-01-13
- Date posted
- 2012-01-24
- Date terminated
- 2012-12-03
- Location
- Ann Arbor, MI