Roche VALP2, Online TDM Valproic acid , for use with the Roche/Hitachi Cobas c 501 analyzer; Catalog No.04491041190.
FDA device recall Z-0920-2007 · posted 2007-06-09 · Terminated
Recall details
- Recalling firm
- Roche Diagnostics Corp.
- Reason for recall
- Probe Wash: Reported results may have a positive bias because inadequate analyzer probe wash instructions may result in carry-over from the following assays: Creatinine Kinase L and Creatine Kinase-MBL.
- Action taken
- Consignees were issued a recall letter dated 4/30/07 instructing them to perform a R2 reagent probe wash when using CKL and CKMBL with the vancomycin or valproic acid assays.
- Root cause
- Other
- Status
- Terminated
- Product code
- LEG
- Quantity affected
- 992
- Distribution
- Nationwide.
- Date initiated
- 2007-04-30
- Date posted
- 2007-06-09
- Date terminated
- 2007-12-17
- Location
- Indianapolis, IN