Atlas FDA

Roche Online TDM Tobramycin, for use on Roche/Hitachi analyzers 917 and Modular P; catalog no. 0464250319011190. (Note:

FDA device recall Z-0918-2007 · posted 2007-06-09 · Terminated

Recall details

Recalling firm
Roche Diagnostics Corp.
Reason for recall
Reported results may have a positive bias of up to 113% because inadequate cleaning instructions may result in carry-over of the following reagents: Online TDM carbamazepine, Online TDM gentamycin, Online TDM phenytoin, Online DAT cocaine II, Online DAT methadone II, Online DAT opiates II and Online DAT THC II.
Action taken
Foreign affiliates were notified via email letter dated 4/5/07 requesting that to notify their customers of the need to implement a special wash cycle to prevent reagent carry over.
Root cause
Other
Status
Terminated
Product code
KLB
Quantity affected
474 total (Z-0917-0918-2007)
Distribution
International Distribution ONLY-Australia, Canada and Germany.
Date initiated
2007-04-05
Date posted
2007-06-09
Date terminated
2008-04-30
Location
Indianapolis, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ONLINE TDM TOBRAMYCIN (K060853)Roche Diagnostics Corp.2006-06-09