Atlas FDA

Roche/Hitachi Glucose Hexokinase Reagent for use on the Hitachi Modular System; Catalog # 1929542 (11929542216). In vitr

FDA device recall Z-0899-06 · posted 2006-05-25 · Terminated

Recall details

Recalling firm
Roche Diagnostics Corp.
Reason for recall
A contaminant will change the pH of the R2 bottle, causing deterioration of the enzyme, and resulting in incorrect or no results being reported.
Action taken
A recall notification letter dated 4/12/06 was sent to each customer instructing them to discontinue use of this lot of product.
Root cause
Other
Status
Terminated
Product code
CFR
Quantity affected
269 kits
Distribution
Nationwide.
Date initiated
2006-04-12
Date posted
2006-05-25
Date terminated
2006-10-16
Location
Indianapolis, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
HITACHI 704 (K854025)Boehringer Mannheim Corp.1985-10-28