Roche Diagnostics NAPA2, N-Acetyl-Procainamide, COBAS INTEGRA, Roche Diagnostics, Indianapolis, IN.; 20737852322. An in-
FDA device recall Z-0012-2009 · posted 2008-10-09 · Terminated
Recall details
- Recalling firm
- Roche Diagnostics Corp.
- Reason for recall
- Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader.
- Action taken
- Consignees were notified via an Urgent Medical Device correction letter dated 6/11/08 informing users of the problem and providing them with work-around instructions. The instructions suggests 1) inserting the cassette rack slowly; 2) using mainly reagent slots A, B, or C; 3) tilting the rack slightly clockwise so the barcode scanner does not read the barcode at an exact 90 degree angle; and 4) confirming the rack is properly seated on the track and displayed on the Cassette Status screen. If the cassette barcode does not register after following the Work-Around instructions, then customers are required to contact Diagnostics Technical Support.
- Root cause
- Labeling Change Control
- Status
- Terminated
- Product code
- CGX
- Quantity affected
- 2
- Distribution
- Nationwide Distribution
- Date initiated
- 2008-06-11
- Date posted
- 2008-10-09
- Date terminated
- 2009-09-22
- Location
- Indianapolis, IN