Roche brand AccuData GTS/GTS Plus System Replacement Base/Module; Catalog number 3426050, material number 03426050001.

FDA device recall Z-0883-05 · posted 2005-06-11 · Terminated

Recall details

Recalling firm
Roche Diagnostics Corp.
Reason for recall
If foil wrapped packages or other metallic objects are lodged behind the supply drawer, an electrical short may result and high heat generated; this could result in plastic melting and battery leakage for customers using 'C' cell batteries to power their base and could render the system inoperable (system won't turn on) for customers using the A/C adaptor without batteries.
Action taken
Consignees were notified via letter dated 4/18/05, which included instructions on how to inspect their unit for damage, and how to prevent future damage.
Root cause
Other
Status
Terminated
Product code
CGA
Distribution
Nationwide and Guam.
Date initiated
2005-04-18
Date posted
2005-06-11
Date terminated
2006-03-07
Location
Indianapolis, IN