RoboCouch Patient Support System, a component of the CyberKnife Robotic Radiosurgery System. Model number 025007, manufa

FDA device recall Z-0856-2009 · posted 2009-01-14 · Terminated

Recall details

Recalling firm
Accuray Inc
Reason for recall
Product may not be tensioned properly, potentially causing unexpected rotation or descent.
Action taken
All affected sites were notified of the potential problem via Customer Advisory Notice dated 30 October 2008. The letter states required actions that the customer should follow at the beginning of each treatment day. The tension of the belt drive and the support bolts in question will be inspected by Accuray Field Service personnel at the affected facilities. In addition, the preventive maintenance procedure will be updated to include regular inspection and replacement of the belt-drive. Any questions or concerns regarding this issue should be directed to Accuray Customer Support at 1-877-668-8667 or 1-408-716-4700 or customersupport@accuray.com.
Root cause
Process control
Status
Terminated
Product code
JAI
Quantity affected
13 units
Distribution
Worldwide Distribution --- including USA and countries of France and Turkey.
Date initiated
2008-10-30
Date posted
2009-01-14
Date terminated
2009-09-10
Location
Sunnyvale, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ROBOCOUCH PATIENT SUPPORT SYSTEM (K042146)Accuray Incorporated2004-09-17