RoboCouch Patient Positioning System Product Usage: is intended for use in the support and positioning of a patient duri

FDA device recall Z-1155-2013 · posted 2013-04-20 · Terminated

Recall details

Recalling firm
Accuray Incorporated
Reason for recall
A5 gearboxes responsible for pitch (head up and head down) and roll (patient left and right) motions of the RoboCouch manipulator may be defective.
Action taken
Accuray issued an Urgent Advisory Notification letter dated March 27, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated that Field Service Engineers will be replacing the affected product. Questions and concerns are to be directed to Accuray Customer Support at 1-877-668-8667
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JAI
Quantity affected
16 devices are subject to correction.
Distribution
Worldwide Distribution - USA Nationwide including the states of: OH, MO. OK, CA and VA and the countries of: Finland, France, Luxembourg, Poland, Russia, Turkey and Japan.
Date initiated
2013-03-27
Date posted
2013-04-20
Date terminated
2013-08-20