Ring Transjugular Intrahepatic Access Set Product Usage: The Ring Transjugular Intrahepatic Access Sets are intended for
FDA device recall Z-2100-2015 · posted 2015-07-18 · Terminated
Recall details
- Recalling firm
- Cook Inc.
- Reason for recall
- Cook Inc. is initiating a voluntary recall of the Ring Transjugular Intrahepatic Access Set and the Flexor RTPS Introduce Sheath due to complaints of the dilators being too long.
- Action taken
- Written recall communication letters were shipped via FedEx on June 05, 2015 to all US Consignees. Customers were asked to complete and return response forms and recalled products. Recalled devices will be held for final disposition and destruction. Customers wree asked to forward notices if recalled devices were further distributed. Effectivenes checks will be conducted by mail, phone and visits. Customers may contact Cook Medical Relations (800) 457-4500 or 1 (812)- 339-2235 Monday through Friday from 7:30 AM to 5:00 PM or e-mail at CustomerrelationsNA@cook medical.com.
- Root cause
- Component change control
- Status
- Terminated
- Product code
- GBZ
- Quantity affected
- 661
- Distribution
- Worldwide Distribution - US Nationwide in the state of: AR, AZ, CA, CO, CT, DC, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SD, TN, TX, UT, VA, WA, WI, WV and t
- Date initiated
- 2015-05-28
- Date posted
- 2015-07-18
- Date terminated
- 2017-05-30
- Location
- Bloomington, IN