Atlas FDA

Resolute Integrity Zotarolimus-eluting Coronary Stent: Stent ID 3.0 mm; Stent length 34 mm; Rx, Sterile. Mfg in Medtroni

FDA device recall Z-0154-2014 · posted 2013-11-05 · Terminated

Recall details

Recalling firm
Medtronic Vascular, Inc.
Reason for recall
A small number of units in lot number 0006573585 of the Resolute Integrity Zotarolimus-Eluting Coronary Sten System may contain a compliance chart incorrectly referencing a 2.25 mm diameter instead of the correct 3.0mm.
Action taken
Urgent Medical Device Recall Letters were sent on October 7, 2013 via UPS 3-day service.
Root cause
Process control
Status
Terminated
Product code
NIQ
Quantity affected
201 units
Distribution
Nationwide Distribution.
Date initiated
2013-10-07
Date posted
2013-11-05
Date terminated
2014-02-11
Location
Santa Rosa, CA

Monitor recalls programmatically