Atlas FDA

Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory s

FDA device recall Z-2735-2026 · posted 2026-07-28 · Open, Classified

Recall details

Recalling firm
ResMed Ltd.
Reason for recall
Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.
Action taken
ResMed Urgent Medical Device Correction Astral 100/150 Ventilators Issue Identified: On June 25, 2026, ResMed notified customers via Salesforce of an urgent medical device correction involving a subset of Astral 100 and Astral 150 ventilators. An internal supercapacitor may leak electrolyte over time, potentially damaging the printed circuit board assembly (PCBA) and causing the ventilator to enter a fail-safe state. Interim Guidance: Due to limited PCBA replacement availability, immediate correction of all affected units is not feasible. Physicians and patients are advised to follow Astral 100/150 User and Clinical Guides, including recommendations on monitoring, emergency preparedness, caregiver training, and backup ventilation. Required Actions for Healthcare Providers and Distributors: Acknowledge: Complete and return the acknowledgment form by July 31, 2026. Notify: Distribute the correction notice and accompanying letters to all relevant healthcare professionals, patients, and caregivers immediately. Reinforce Safety Measures: Ensure ventilator-dependent patients are monitored, caregivers are trained in alarm response and emergency procedures, and backup ventilation equipment is functional and readily available. Device Management: Do not remove an affected ventilator from use unless an alternative is immediately available. Replace inoperable units promptly; patients should continue therapy unless directed otherwise by their clinician. Risk Assessment: Evaluate affected patients by clinical risk category (Tier 1, 2, or 3) per Appendix A. Arrange Inspection: Schedule ventilator inspections per the Phase 1 Inspection and Correction Strategy in Appendix A, returning units to an authorized service center as needed. Contact: Reach your local ResMed representative or visit www.resmed.com/contact for questions.
Root cause
Component design/selection
Status
Open, Classified
Product code
CBK
Quantity affected
168,069 devices globally
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY,
Date initiated
2026-06-25
Date posted
2026-07-28
Location
Bella Vista, Australia

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Astral 100/150 (K172875)Resmed, Ltd.2018-04-26
Astral 100/150 (K152068)Resmed, Ltd.2016-05-13