Atlas FDA

Replacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010

FDA device recall Z-1565-2023 · posted 2023-05-16 · Open, Classified

Recall details

Recalling firm
ICU Medical Inc
Reason for recall
Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.
Action taken
ICU medical issued an URGENT: MEDICAL DEVICE CORRECTION notice on 03/22/2023 to its direct and indirect consignees by mail. The notice explained the issue and the hazard and requested the following: Clinical Users: Whenever possible, keep the pump plugged into AC power. Before disconnecting the pump from AC power, please ensure that the battery is fully charged. Closely monitor the Battery Status Indicator while the pump is disconnected from AC power. Additionally, have a backup pump available when infusing critical medications. Biomedical Engineering: You may replace affected batteries with a new CSB battery until corrected batteries are available. Please do not use a replacement battery with corroded battery terminals. 1. Identify all affected batteries in your possession and ensure all users or potential users of these pumps are immediately made aware of this notification and proposed mitigations. 2. Complete and return the attached Response Form to icumedical5967@sedgwick.com within ten days of receipt to acknowledge your understanding of this notification. 3. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to icumedical5967@sedgwick.com.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
FRN
Quantity affected
28656 units
Distribution
worldwide
Date initiated
2023-03-22
Date posted
2023-05-16
Location
Lake Forest, IL

Monitor recalls programmatically