REF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 mm 64 mL/min , Rx Only

FDA device recall Z-0025-2024 · posted 2023-10-03 · Open, Classified

Recall details

Recalling firm
Becton Dickinson Infusion Therapy Systems Inc.
Reason for recall
There is the potential that the safety shield may not properly engage on IV catheters. The needle can pull through the safety shield and disconnect from the needle safety shield leaving the contaminated needle point exposed.
Action taken
On August 29, 2023, Becton Dickinson issued an "Urgent Medical Device Recall" notification via. FedEx. Becton Dickinson asked consignees to take the following actions: 1. Immediately review your inventory for Catalog Number 386862, Lot number 2238464. Destroy all unused product subject to the recall following your institution's process for destruction. 2. For indwelling catheters, no further action is needed. 3. If a defective product was used on a patient and the used needle was disposed of without the clinician suffering a needlestick injury, no further action is necessary to further protect the patient or the clinician. 4. Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 5. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement. 6. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch, Phone: 1-800-FDA 1088 (1-800-332-1088), Mail: MedWatch, HF-2, FDA, 5600 Fisher.s Lane, Rockville, MD 20852-978
Root cause
Process change control
Status
Open, Classified
Product code
FOZ
Quantity affected
98,000 units
Distribution
US Nationwide distribution in the states of CA, FL, GA, MA, MN, MO, MS, NJ, OH, TX.
Date initiated
2023-08-29
Date posted
2023-10-03
Location
Sandy, UT