RECLAIM DISTAL TAPERED, Angled Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedi
FDA device recall Z-2083-2014 · posted 2014-07-21 · Terminated
Recall details
- Recalling firm
- DePuy Orthopaedics, Inc.
- Reason for recall
- The product can be difficult to remove from the Distal Stem both out of the package and after proximal reaming.
- Action taken
- DePuy Synthes sent an Urgent Information - Device Correction Notice letter dated June 24, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: 1. Immediate Remedy: Discard the RECLAIM Taper Sleeve Protector instrument before stem implantation. 2. Changes to the literature and IFU: Documents are revised as follows: " RECLAIM Revision Hip Systems Surgical Technique # DSUS/JRC/0614/0200 pages 11, 12, 13, 16, 30, 34, 35 (figure numbering from pages 11 to 27 were also revised): Removed information regarding taper sleeve protector. " RECLAIM Revision Hip Systems Surgical Technique FAQ # DSUS/JRC/0614/0202 pages 1 and 2: Removed information regarding taper sleeve protector. " RECLAIM Revision Hip Systems IFU #090200799 REV E Page 5: Verbiage will be revised and implemented as soon as possible, depending on regulatory approvals. 3. Change: The company consulted with design surgeons and developed a strategy to eliminate the taper sleeve protector from the system. As a result of removing the taper sleeve protector, the company is evaluating whether there is a need for a protective component to be used during the distribution Customers with questions were instructed to call 574-371-4917. For questions regarding this recall call 574-371-4577.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- KWA
- Quantity affected
- 10,467 units
- Distribution
- Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, S
- Date initiated
- 2014-06-03
- Date posted
- 2014-07-21
- Date terminated
- 2017-05-17
- Location
- Warsaw, IN