RaySearch RayStation 2.5, 3.0, 3.5 and 4.0; Radiation Therapy Treatment Product Usage: RayStation is a software system d
FDA device recall Z-1679-2015 · posted 2015-05-27 · Terminated
Recall details
- Recalling firm
- RAYSEARCH LABORATORIES AB
- Reason for recall
- RaySearch Laboratories AB became aware of an issue with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing.
- Action taken
- RaySearch sent a Device Correction letter dated May 8, 2015 via e-mailed to their customers. The letter identified the affected product, problem, actions to be taken by users and contact information.
- Root cause
- Software Design Change
- Status
- Terminated
- Product code
- MUJ
- Quantity affected
- Domestic: 539 units
- Distribution
- US Nationwide Distribution
- Date initiated
- 2015-05-08
- Date posted
- 2015-05-27
- Date terminated
- 2017-04-11
- Location
- Stockholm, Sweden