RaySearch RayStation 2.5, 3.0, 3.5 and 4.0; Radiation Therapy Treatment Product Usage: RayStation is a software system d

FDA device recall Z-1679-2015 · posted 2015-05-27 · Terminated

Recall details

Recalling firm
RAYSEARCH LABORATORIES AB
Reason for recall
RaySearch Laboratories AB became aware of an issue with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing.
Action taken
RaySearch sent a Device Correction letter dated May 8, 2015 via e-mailed to their customers. The letter identified the affected product, problem, actions to be taken by users and contact information.
Root cause
Software Design Change
Status
Terminated
Product code
MUJ
Quantity affected
Domestic: 539 units
Distribution
US Nationwide Distribution
Date initiated
2015-05-08
Date posted
2015-05-27
Date terminated
2017-04-11
Location
Stockholm, Sweden