Atlas FDA

Rascal 600F and 600T scooters, unspecified power wheelchair.

FDA device recall Z-0903-06 · posted 2006-05-24 · Terminated

Recall details

Recalling firm
Electric Mobility Corp
Reason for recall
The four seat mast mounting screws can loosen, causing seat to wobble. The chair seat actuator and portions of the seat can malfunction and/or break. Firm is aware of problems with the seat and injuries sustained from the powered wheelchair/scooter seat malfunction. Firm has not initiated recall, claims it is a Warranty Advisory.
Action taken
At this point, the firm has committed to send out a Warranty advisory. No recall has been initiated.
Root cause
Other
Status
Terminated
Product code
ITI
Quantity affected
12,220 chairs
Distribution
To customers nationwide.
Date initiated
2006-04-17
Date posted
2006-05-24
Date terminated
2014-04-08
Location
Sewell, NJ