RAPIDPOINT 500e Blood Gas System
FDA device recall Z-0422-2024 · posted 2023-12-01 · Open, Classified
Recall details
- Recalling firm
- Siemens Healthcare Diagnostics Inc
- Reason for recall
- There are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The erroneous sodium levels with positive bias may lead to unrecognized hyponatremia and/or the inappropriate treatment of hypernatremia may lead to iatrogenic hyponatremia, resulted in a delay in patient diagnosis and optimal patient management.
- Action taken
- The firm issued an Urgent Medical Device Correction notice on 9/26/2023 to customers asking that customers review the notice with the facility's Medical Director to determine the appropriate course of action for previously generated results, complete and return the Field Correction Effectiveness Check Form within 30 days, forward this notice to customers if your firm is a distributor, and retain the notification in your system's Operator's Guide and forward to those who may have received product. The Field Correction Effectiveness Check Form is to be returned via email to uscctsfcaecfax.team@siemens-healthineers.com or by fax to 312-275-7795. Questions are to be directed to your local Siemens Healthineers Technical Support Representative.
- Root cause
- Under Investigation by firm
- Status
- Open, Classified
- Product code
- JGS
- Quantity affected
- 8,749 units
- Distribution
- Domestic: Nationwide Distribution. Foreign: Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswa
- Date initiated
- 2023-09-26
- Date posted
- 2023-12-01
- Location
- Norwood, MA