RAPIDPoint 500 Measurement Clinical Chemistry System RAPIDPoint 500 Measurement Cartridge Lactate (750 Tests) Part numbe
FDA device recall Z-2079-2014 · posted 2014-07-18 · Terminated
Recall details
- Recalling firm
- Siemens Healthcare Diagnostics Inc
- Reason for recall
- RAPIDPoint 500 Measurement Cartridges may have an error code one or more electrolyte parameters (e.g., Na+, Ca++, K+, Cl-)
- Action taken
- Siemens Healthcare Diagnostics issued an Urgent Field Safety Notice by e-mail to all affected Siemens Healthcare Diagnostics Regional offices on May 27, 2014 both in the United States and elsewhere for communication with affected customers. This notice informs customers of the issue with the RAPIDPoint 500 Measurement cartridges. Questions contact: Siemens Customer Care Center or your local Siemens technical support representative at 877-229-3711. What steps can be taken to mitigate the issue: " Install a new RAPIDPoint 500 Measurement Cartridge. or " Install a 405 Measurement Cartridge - The RAPIDPoint 405 cartridge is available in four sizes: Since the Lactate parameter is not available on the RAPIDPoint 405 cartridge, however, you will need to arrange to get your lactate results from other instruments in your central laboratory. or " Use an alternative system to obtain your electrolyte values.
- Root cause
- Process design
- Status
- Terminated
- Product code
- KHP
- Quantity affected
- 20,507 cartridges
- Distribution
- Nationwide Foreign: U.A.E. Albania Austria Australia Bosnia Herzeg. Belgium Bulgaria Bahrain Brazil Canada Switzerland P.R. China Colombia Czech Republic Germany Denmark Algeria Estonia Egypt Spain Fi
- Date initiated
- 2014-05-23
- Date posted
- 2014-07-18
- Date terminated
- 2015-06-10
- Location
- Norwood, MA