RAPIDPoint 500 Measurement Clinical Chemistry System RAPIDPoint 500 Measurement Cartridge Lactate (750 Tests) Part numbe

FDA device recall Z-2079-2014 · posted 2014-07-18 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics Inc
Reason for recall
RAPIDPoint 500 Measurement Cartridges may have an error code one or more electrolyte parameters (e.g., Na+, Ca++, K+, Cl-)
Action taken
Siemens Healthcare Diagnostics issued an Urgent Field Safety Notice by e-mail to all affected Siemens Healthcare Diagnostics Regional offices on May 27, 2014 both in the United States and elsewhere for communication with affected customers. This notice informs customers of the issue with the RAPIDPoint 500 Measurement cartridges. Questions contact: Siemens Customer Care Center or your local Siemens technical support representative at 877-229-3711. What steps can be taken to mitigate the issue: " Install a new RAPIDPoint 500 Measurement Cartridge. or " Install a 405 Measurement Cartridge - The RAPIDPoint 405 cartridge is available in four sizes: Since the Lactate parameter is not available on the RAPIDPoint 405 cartridge, however, you will need to arrange to get your lactate results from other instruments in your central laboratory. or " Use an alternative system to obtain your electrolyte values.
Root cause
Process design
Status
Terminated
Product code
KHP
Quantity affected
20,507 cartridges
Distribution
Nationwide Foreign: U.A.E. Albania Austria Australia Bosnia Herzeg. Belgium Bulgaria Bahrain Brazil Canada Switzerland P.R. China Colombia Czech Republic Germany Denmark Algeria Estonia Egypt Spain Fi
Date initiated
2014-05-23
Date posted
2014-07-18
Date terminated
2015-06-10
Location
Norwood, MA