RAPIDPoint 500 Blood Gas Analyzer, Siemens Material Numbers (SMN):10492730, 10696855, 10696857, 10697306 This system tes

FDA device recall Z-2479-2015 · posted 2015-08-28 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics Inc
Reason for recall
When both ports (Serial and Ethernet) are configured to transmit data, the data stream from one port could potentially affect the data stream from the other port causing the data management system to include either: duplicate data, missing data, or data from a different patient.
Action taken
Siemens Healthcare issued Urgent Field Safety Notice sent by e-mail to all affected Siemens Healthcare Diagnostics Regional offices on July 28, 2015, both in the United States and outside of the United States, for communication with affected customers. This notice informs customers of the issue with the RAPIDPoint 500 dual port LIS transmission feature. Users are informed to: If the system is currently set up with dual LIS transmission enabled, one of the ports (either Serial or Ethernet) needs to be disabled. Instruction provided with the notification. Customer to complete Field Effectiveness Check Form to confirm receipt of the notice. The resolution of this issue will be implemented in the next software release for the RAPIDPoint 500 system. If you have any questions, contact Siemens Customer Care Center or your local Siemens technical support representative. Please fax this completed Customer Care form to (312) 275-7795.
Root cause
Software design
Status
Terminated
Product code
KHP
Quantity affected
5241 units
Distribution
Worldwide Distribution - US (nationwide) Distribution and to the countries of : Albania, Algeria, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Bulgaria,Canada, Chi
Date initiated
2015-07-28
Date posted
2015-08-28
Date terminated
2017-02-17
Location
Norwood, MA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
RAPIDPOINT 500 BLOOD GAS ANALYER (K113216)Siemens Healthcare Diagnostics2012-05-03
RAPIDPOINT 405 NEONATAL BILIRUBIN (K110277)Siemens Healthcare Diagnostics2011-08-15
BAYER RAPIDPOINT 405 SYSTEM (K020616)Bayer Diagnostics Corp.2002-07-17
BAYER RAPIDPOINT 400 SYSTEM (K002738)Bayer Corp.2000-12-19