RAPIDPoint 500 Blood Gas Analyzer, Siemens Material Numbers (SMN):10492730, 10696855, 10696857, 10697306 This system tes
FDA device recall Z-2479-2015 · posted 2015-08-28 · Terminated
Recall details
- Recalling firm
- Siemens Healthcare Diagnostics Inc
- Reason for recall
- When both ports (Serial and Ethernet) are configured to transmit data, the data stream from one port could potentially affect the data stream from the other port causing the data management system to include either: duplicate data, missing data, or data from a different patient.
- Action taken
- Siemens Healthcare issued Urgent Field Safety Notice sent by e-mail to all affected Siemens Healthcare Diagnostics Regional offices on July 28, 2015, both in the United States and outside of the United States, for communication with affected customers. This notice informs customers of the issue with the RAPIDPoint 500 dual port LIS transmission feature. Users are informed to: If the system is currently set up with dual LIS transmission enabled, one of the ports (either Serial or Ethernet) needs to be disabled. Instruction provided with the notification. Customer to complete Field Effectiveness Check Form to confirm receipt of the notice. The resolution of this issue will be implemented in the next software release for the RAPIDPoint 500 system. If you have any questions, contact Siemens Customer Care Center or your local Siemens technical support representative. Please fax this completed Customer Care form to (312) 275-7795.
- Root cause
- Software design
- Status
- Terminated
- Product code
- KHP
- Quantity affected
- 5241 units
- Distribution
- Worldwide Distribution - US (nationwide) Distribution and to the countries of : Albania, Algeria, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Bulgaria,Canada, Chi
- Date initiated
- 2015-07-28
- Date posted
- 2015-08-28
- Date terminated
- 2017-02-17
- Location
- Norwood, MA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| RAPIDPOINT 500 BLOOD GAS ANALYER (K113216) | Siemens Healthcare Diagnostics | 2012-05-03 |
| RAPIDPOINT 405 NEONATAL BILIRUBIN (K110277) | Siemens Healthcare Diagnostics | 2011-08-15 |
| BAYER RAPIDPOINT 405 SYSTEM (K020616) | Bayer Diagnostics Corp. | 2002-07-17 |
| BAYER RAPIDPOINT 400 SYSTEM (K002738) | Bayer Corp. | 2000-12-19 |