Rapid Response HUT (Head Up Tilt) Table. Model 1117. Powered electrophysiology tilt table with single drop section. Tabl

FDA device recall Z-1974-2011 · posted 2011-04-13 · Terminated

Recall details

Recalling firm
Medical Positioning Inc
Reason for recall
The tilt tables work intermittently and some have become stuck in an inclined or reclined position during procedures.
Action taken
An undated Urgent - Voluntary Medical Device Product Correction letter was issued on 2/1/2010 by certified mail with return receipt requested. The letter identified the model and serial numbers of the affected product. The consignees were advised to stop using the table immediately and remove it from service until further notice. Customers were to contact Mr. Joe Hurtig at 1-800-593-3246 or jhurtig@medicalpositioning.com to confirm receipt of the notice and to provide the firm with the correct contact information. Individualized correction packages including replacement parts and instructions were sent to each consignee. The instructions consisted of (1) Rapid Response HUT Motor Replacement Instructions; (2) Control Box Replacement Instructions; and (3) ILD External Power Supply Removal Instructions.
Root cause
Other
Status
Terminated
Product code
IXR
Quantity affected
58 total
Distribution
Worldwide Distribution -- USA, including Puerto Rico, and Canada.
Date initiated
2010-02-01
Date posted
2011-04-13
Date terminated
2011-06-15
Location
Kansas City, MO