Rapid Response H.U.T. (Head Up Tilt) Table. Model 1017. Powered electrophysiology tilt table. Table can tilt -15 degrees
FDA device recall Z-1973-2011 · posted 2011-04-13 · Terminated
Recall details
- Recalling firm
- Medical Positioning Inc
- Reason for recall
- The tilt tables work intermittently and some have become stuck in an inclined or reclined position during procedures.
- Action taken
- An undated Urgent - Voluntary Medical Device Product Correction letter was issued on 2/1/2010 by certified mail with return receipt requested. The letter identified the model and serial numbers of the affected product. The consignees were advised to stop using the table immediately and remove it from service until further notice. Customers were to contact Mr. Joe Hurtig at 1-800-593-3246 or jhurtig@medicalpositioning.com to confirm receipt of the notice and to provide the firm with the correct contact information. Individualized correction packages including replacement parts and instructions were sent to each consignee. The instructions consisted of (1) Rapid Response HUT Motor Replacement Instructions; (2) Control Box Replacement Instructions; and (3) ILD External Power Supply Removal Instructions.
- Root cause
- Other
- Status
- Terminated
- Product code
- IXR
- Quantity affected
- 58 total
- Distribution
- Worldwide Distribution -- USA, including Puerto Rico, and Canada.
- Date initiated
- 2010-02-01
- Date posted
- 2011-04-13
- Date terminated
- 2011-06-15
- Location
- Kansas City, MO