rainbow Reusable Sensors. RAINBOW DCI-DC3, SpO2/SpCO/SpMet, Part No. 2201; RAINBOW DCI-dc8, SpO2/SpCO/SpMet, Part No. 24
FDA device recall Z-0338-2016 · posted 2015-11-25 · Terminated
Recall details
- Recalling firm
- Masimo Corporation
- Reason for recall
- These sensors were manufactured with incompatible configurations. This could result in sensors that will provide either no readings or inaccurate readings for the SpCO and SpMet parameters.
- Action taken
- Masimo sent an Urgent Product Recall letter on October 28, 2015, to all affected customers to inform them that Masimo has identified specific lots of sensors that were manufactured with incompatiable configurations. The letter provides the customers with the problems identified and the actions to be taken. Customers are instructed to complete the tracking/verification form and email to Masimo. Customers with questions were instructed to contact Masimo Technical Services at http://Service.Masimo.com. For questions regarding this recall call 949-297-7000.
- Root cause
- Finished device change control
- Status
- Terminated
- Product code
- DQA
- Quantity affected
- 3476 units total (3322 units in US)
- Distribution
- Worldwide Distribution - US (nationwide) and Internationally to Australia, Costa Rica, France, Germany, Kuwait, Martinique, Netherlands, Norway, South Korea, and Turkey.
- Date initiated
- 2015-10-28
- Date posted
- 2015-11-25
- Date terminated
- 2016-05-18