Radiohead Access Point 1.4 GHz Radiohead Access Point Firmware versions: C.00.04 & C.00.05 Part Number: 989803171211 Pro

FDA device recall Z-1780-2013 · posted 2013-07-20 · Terminated

Recall details

Recalling firm
Philips Healthcare Inc.
Reason for recall
IntelliVue Smart-hopping 1.4 GHz Access Point may experience unexpected signal loss of several minutes impacting wireless connectivity
Action taken
Philips sent an Urgent Medical Device Correction letter dated July 2013 to affected customers. The letter informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer. Philips will perform a firmware upgrade (version C.00..07 for all affected customers at no cost. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice. If you need any further information or support concerning this issue, please contact Philips Customer Support at 1-800-722-9377
Root cause
Software Design Change
Status
Terminated
Product code
MHX
Quantity affected
20105 units
Distribution
USA Nationwide Distribution
Date initiated
2013-07-08
Date posted
2013-07-20
Date terminated
2017-05-17
Location
Andover, MA