RadiForce RX850, 8MP Color LCD Monitor, Part Code No. 0FTD1958A2, UPC 690592037180. This product is intended to be used
FDA device recall Z-2089-2015 · posted 2015-07-16 · Terminated
Recall details
- Recalling firm
- Eizo Corporation
- Reason for recall
- EIZO is recalling the RadiForce LCD Monitors could display abnormally while connecting to particular graphics boards with DisplayPort.
- Action taken
- EIZO sent An Urgent Field Safety Notice dated July 7, 2015, to all customers who purchased the RadiForce RX440, RX650, RX850 and GX540. The notice informs customers that the RadiForce monitors could display abornormally in 6-bit (64 gradations) instead of normal 8-bit (256 gradations) or 10-bit (1024 gradations) while connecting to particular graphics boards with DisplayPort. The notice provides the customers with the actions to be taken and contact information (1-800-800-5202, M-F, 8am-4pm, PST, us.feedback@eizo.com, 5710 Warland Drive, Cypress, CA 90630).
- Root cause
- Device Design
- Status
- Terminated
- Product code
- PGY
- Quantity affected
- 2 units
- Distribution
- Nationwide Distribution including OH, FL, IL, TX, OH, NY, AZ, and NJ.
- Date initiated
- 2015-07-07
- Date posted
- 2015-07-16
- Date terminated
- 2016-11-30