Atlas FDA

RadiForce RX850, 8MP Color LCD Monitor, Part Code No. 0FTD1958A2, UPC 690592037180. This product is intended to be used

FDA device recall Z-2089-2015 · posted 2015-07-16 · Terminated

Recall details

Recalling firm
Eizo Corporation
Reason for recall
EIZO is recalling the RadiForce LCD Monitors could display abnormally while connecting to particular graphics boards with DisplayPort.
Action taken
EIZO sent An Urgent Field Safety Notice dated July 7, 2015, to all customers who purchased the RadiForce RX440, RX650, RX850 and GX540. The notice informs customers that the RadiForce monitors could display abornormally in 6-bit (64 gradations) instead of normal 8-bit (256 gradations) or 10-bit (1024 gradations) while connecting to particular graphics boards with DisplayPort. The notice provides the customers with the actions to be taken and contact information (1-800-800-5202, M-F, 8am-4pm, PST, us.feedback@eizo.com, 5710 Warland Drive, Cypress, CA 90630).
Root cause
Device Design
Status
Terminated
Product code
PGY
Quantity affected
2 units
Distribution
Nationwide Distribution including OH, FL, IL, TX, OH, NY, AZ, and NJ.
Date initiated
2015-07-07
Date posted
2015-07-16
Date terminated
2016-11-30